Molecular Magnetic Resonance Imaging of Coronary Thrombosis and Pulmonary Emboli With a Novel Fibrin-Targeted Contrast Agent
http://www.100md.com 循环学杂志 2005年第3期
     the Department of Diagnostic Radiology, Aachen Technical University, Aachen, Germany (E.S., A.B., M.K., R.W.G.)

    EPIX Pharmaceuticals, Cambridge, Mass (A.J.W., E.C.P., P.B.G.)

    Departments of Medicine (Cardiovascular Division) (R.M.B., W.J.M.) and Radiology (W.J.M.)

    Beth Israel Deaconess Medical Center, and Harvard Medical School, Boston, Mass.

    Reprint requests to Elmar Spuentrup, MD, Department of Diagnostic Radiology, University Hospital, Aachen Technical University, Pauwelsstrasse 30, 52057 Aachen, Germany. E-mail spuenti@rad.rwth-aachen.de

    Abstract

    Background— The differential diagnosis of acute chest pain is challenging, especially in patients with normal ECG findings, and may include coronary thrombosis or pulmonary emboli. The aim of this study was to investigate the novel fibrin-specific contrast agent EP-2104R for molecular targeted MR imaging of coronary thrombosis and pulmonary emboli.

    Methods and Results— Fresh clots were engineered ex vivo from human blood and delivered in the lungs and coronary arteries of 7 swine. Subsequent molecular MR imaging was performed with a navigator-gated free-breathing and cardiac-triggered 3D inversion-recovery black-blood gradient-echo sequence before and after systemic administration of 7.5 μmol/kg EP-2104R. Two swine served as the control group. MR images were analyzed by 2 investigators, and contrast-to-noise ratio and gadolinium concentration in the clots were assessed. Before contrast media application, no thrombi were visible. After contrast administration, all 32 pulmonary emboli, 3 emboli in the right heart, and 5 coronary thrombi were selectively visualized as white spots with a mean contrast-to-noise ratio of 32±19. The average gadolinium concentration from all 3 types of thrombi was 144±79 μmol/L.

    Conclusions— Molecular MR imaging with the fibrin-targeted contrast-agent EP-2104R allows selective visualization of acute coronary, cardiac, and pulmonary thrombi.

    Key Words: contrast media ; coronary vessels ; fibrin ; magnetic resonance imaging ; thrombosis

    Introduction

    Patients with chest pain are frequently sent to the emergency room for diagnostic evaluation and treatment.1,2 In 30% of these patients, the primary cause is related to cardiac disease, and in only 15% is myocardial infarction with typical ECG changes seen. Hence, further differential diagnoses such as acute coronary syndromes without myocardial necrosis (unstable angina), non–ST-elevation infarction, or pulmonary embolism also have to be taken into account.1,2 These diseases may also be life threatening because of potential subsequent myocardial infarction or circulatory collapse.

    Unstable angina and non–ST-elevation myocardial infarction are believed to be caused by plaque rupture and subsequent coronary thrombus formation with potential risk of vessel occlusion.3–5 Currently, coronary thrombosis can be visualized only invasively by x-ray angiography, intravascular ultrasound, or angioscopy. The decision to use these invasive procedures must depend on the patient-specific total risk, including knowledge of the ischemic cause of the symptoms.6

    Pulmonary embolism may also be a differential diagnosis in patients with chest pain (with or without dyspnea). However, clinical findings are often unspecific.7–9 Furthermore, serum troponin (a marker for myocardial necrosis) can be elevated in patients with pulmonary embolism.10 Thus, a noninvasive imaging modality that allows direct visualization of coronary thrombosis and pulmonary emboli would be of great value.

    In the present study, selective and noninvasive molecular MR imaging of coronary thrombi and pulmonary emboli is demonstrated by means of a fibrin-targeted MR contrast agent (EP-2104R, EPIX Pharmaceuticals).

    Methods

    Subjects

    Noninvasive molecular MR imaging of coronary thrombi and pulmonary emboli was performed in 7 healthy domestic swine (48 to 52 kg) as approved by the German government committee on animal investigations. After premedication with 0.5 mL IM atropine, 0.2 mL IM azaperone/kg body weight, and 0.1 mL ketamine/kg body weight, an aqueous solution of pentobarbital (1:3) was administered intravenously via an ear vein as needed. The animals were intubated, and mechanical ventilation was maintained throughout the entire experiment. A 9F sheath (Cordis) was placed surgically in the right carotid artery, and a 16F sheath (Cordis) was placed in right iliac vein.

    Clot Engineering

    Thrombi were formed ex vivo for injection into the coronary arteries and lungs. For each 1-mL thrombus, 10 mg human fibrinogen, 10 NIH units of thrombin, 25 mmol/L CaCl2, and blood were mixed in a 1-mL syringe and allowed to sit for 30 minutes. A portion of the thrombi (0.2 mL) was used for coronary thrombi delivery. For the pulmonary thrombi, 0.5 mL segments were delivered.

    MR Imaging

    All studies were carried out on a 1.5-T Gyroscan Intera whole-body MR system (Philips Medical Systems; 23 mT/m, 219-μs rise time). A 4-element body-wraparound phased-array coil was used for signal reception. All subjects were examined in supine position.

    Pulmonary MR Imaging

    MR imaging of the lung consisted of a navigator-gated free-breathing and cardiac-triggered inversion-recovery and fat-suppressed 3D black-blood gradient-echo sequence (Figure 1). Sequence parameters include the following: repetition time, 4.0 ms; echo time, 1.3 ms; flip angle, 30°; field of view, 400x400 mm; and matrix, 256x256 reconstructed with a 512x512 matrix to a 0.75x0.75x2-mm voxel size, including zero filling in the z direction. The blood signal was nulled by use of an inversion prepulse, with heart rate–specific inversion times for optimized black-blood imaging.11 A gradient-echo readout with 36 excitations per R-R interval resulted in a relatively brief (145 ms) acquisition window. Data were acquired in late diastole to minimize cardiac motion artifacts. Furthermore, central k-space lines were acquired first to optimize prepulse efficiency and to minimize respiratory motion artifacts. Eighty slices (2 mm) were acquired in the coronal plane.

    Coronary MR Imaging

    For coronary MR imaging, a transverse 3D steady-state free-precession scout scan covering the entire heart was performed first for planning of the subsequent targeted double-oblique coronary MR angiography and coronary thrombus imaging. Coronary bright-blood MR angiography was performed with a previously described navigator-gated free-breathing and cardiac-triggered T2-prepared 3D steady-state free-precession sequence allowing visualization of the anatomy of the coronary artery lumen.12 MR thrombus imaging was also performed by use of a navigator-gated free-breathing and cardiac-triggered inversion-recovery and fat-suppressed 3D black-blood gradient-echo sequence (Figure 1). Spatial resolution of the coronary thrombus scan was increased to 0.6x0.6x1.5 mm while all other parameters were maintained. To account for more extensive motion of the coronary arteries, the end-diastolic acquisition window was reduced to 12 excitations per R-R interval (56-ms end-diastolic acquisition window), allowing further reduction of intrinsic cardiac motion artifacts. Both white-blood coronary MRA and black-blood coronary thrombus imaging were performed in the same double-oblique imaging planes covering twenty-four 1.5-mm-thick slices.

    Two-Dimensional Selective Navigator

    For free-breathing data acquisition, all sequences were equipped with a right hemi-diaphragmatic prospective real-time navigator for respiratory motion-artifact suppression.12 A gating window of 5 mm was used. Because the inversion pulse in the inversion recovery black-blood sequences may reduce navigator performance, the excitation angle of the navigator beam was increased to 45°, and the navigator diameter was set to 50 mm. This allowed high navigator performance, with navigator efficiency always >50%, while maintaining correct navigator position detection. The navigator restore pulse13 was switched off because this pulse may result in a "spin labeling" of the pulmonary blood, potentially reducing the effectiveness of nulling the blood signal in the lungs.

    Fibrin-Targeted Contrast Agent

    The novel fibrin-specific MR contrast agent EP-2104R is composed of a small peptide with 4 gadolinium (Gd)-chelate moieties. EP-1873, a similar compound, has been shown to bind to fibrin without binding to circulating fibrinogen.14,15 EP-2104R (15 mmol/L in saline) was infused over 3 minutes for a dose of 7.5 μmol/kg.

    Pulmonary Multidetector CT Angiography

    Multidetector CT (MDCT) was performed for comparison for its demonstrated ability to detect pulmonary embolism in a swine model.16–18 CT is currently used clinically in patients with suspected pulmonary embolism7,18,19 and is a tomographic imaging modality that allows a more direct comparison to MR imaging compared with projection techniques like x-ray pulmonary angiography. Furthermore, it is very fast to perform, avoiding any major time delay between contrast media injection and autopsy. Sixteen-row MDCT scanning of the lung was performed with 16x0.75-mm collimation (Somatom Sensation, Siemens) with 120-kV tube voltage, 300-mm reconstruction field of view, and 15-mm table feed per rotation after bolus application of 90 mL nonionic contrast material (Ultravist 370, Schering) at a flow rate of 3.5 mL/s. Axial images with 2-mm reconstruction increment and coronal multiplanar reconstructions from 1.0/0.6-mm reconstructions were used as described in literature.18

    Description of Experiments

    For pulmonary embolism, 4 to 8 thrombi per swine were placed in a 12F catheter, and the emboli were delivered via the 16F sheath to the iliac vein by pushing with saline. Coronary thrombi were delivered via a 9F guiding catheter into the left anterior descending (n=3), right coronary (n=1), and left circumflex (n=1) arteries under x-ray guidance. In one control experiment, 7 pulmonary emboli were delivered and imaged without application of extrinsic contrast medium. MR imaging in this pig was repeated after 4 and 6 hours to exclude any signal changes of the clot itself during the experiments. In a second control swine (8 pulmonary emboli), MR imaging was performed after administration of standard extracellular contrast with a clinical dose (0.1 mmol/kg Gd-DTPA, Magnevist, Schering). All pigs were given heparin to avoid additional clotting.

    After clot delivery, the pigs were transferred to the MR unit. Identical MR thrombus imaging sequences of the lungs (coronal slice orientation) and coronary arteries (double-oblique slice orientation along the main axis of the coronary arteries) were performed before contrast media application and repeated 2 hours after systemic application of 7.5 μmol EP2104R/kg via an ear vein. Because scanning time varies, depending on heart rate and navigator gating, the time delay between contrast application and imaging varied slightly between 100 and 135 minutes after contrast media injection. After MR imaging, MDCT was performed; immediately afterward, animals were overdosed with heparin and killed, and clots were removed for assessment of Gd concentration.

    Data Analyses

    MR images were analyzed by 2 radiologists by consensus before MDCT and autopsy. Number and localization of readily visible clots were recorded. These data were compared with findings on MDCT and autopsy. For assessment of interobserver variance, images were again reviewed by a second group of 2 investigators blinded to data of the MDCT and autopsy.

    Signal analyses included determination of contrast-to-noise ratio (CNR) between the clot and surrounding blood pool. The regions of interest were placed manually in the postcontrast images and then copied into the precontrast images. CNR was calculated as follows: CNR equals the clot signal minus the blood signal divided by the SD of noise, which was determined in a region of air outside the chest.

    Gd concentrations were measured by inductively coupled plasma-mass spectrometry. Thrombi were digested overnight with concentrated nitric acid. Digested samples were then diluted with 5% nitric acid containing Tb as an internal standard. The Gd concentration was determined against a calibration curve of the Gd to internal standard ratio versus concentration.

    Statistical Analysis

    CNR values of precontrast and postcontrast images were compared by use of a paired 2-tailed Student t test. A value of P<0.05 was considered significant. Interobserver agreement was assessed through the use of statistics.

    Results

    In 7 pigs (n=7), clot delivery and coronary and pulmonary MR imaging sequences were successfully completed. One additional pig not included in this study died during coronary artery thrombus delivery.

    Before contrast media injection, neither pulmonary emboli nor coronary thrombi were observed (Figures 2 through 4). Two hours after systemic administration of EP2104R, pulmonary emboli (n=32; Figures 2 and 3) and coronary thrombi (n=5; Figure 4) were seen as white spots with a high contrast to the surrounding tissue, independent of the localization of the clots. In addition to the pulmonary emboli, 3 emboli in the right heart (2 in the right chamber, 1 in the right atrium) were observed with a bright signal (Figure 3, example IV). A very high agreement between both groups of radiologists was found (=0.92).

    Coronary thrombi and pulmonary emboli were confirmed by x-ray angiography, MDCT, and autopsy. However, 2 pulmonary emboli located adjacent to the aorta or heart and 2 emboli in the right heart were missed on MDCT, probably because of aggravated motion artifacts on CT images in these regions. However, the existence of these emboli was proven macroscopically.

    CNR measurements yielded a highly significant increase in CNR after contrast media application for lung, right heart, and coronary clots (2±2/4±1/2±1 versus 36±11/56±14/17±6, respectively; P<0.05). The average Gd concentration for lung, chamber, and coronary clots (n=33) was 144±79 μmol/L. The weight and concentration ranges for each group of thrombi are shown in the Table. In the pig that did not receive any contrast medium, there was no Gd in any of the thrombi (n=8). In the pig that received standard extracellular contrast medium (Gd-DTPA), the Gd concentration was 22±16 μmol/L (n=7 clots).

    Gd Concentration

    In both control pigs (after administration of standard extracellular contrast medium also up to 6 hours after clot delivery without any contrast media application), no thrombi could be delineated. Hence, CNR measurements could be performed only in the swine that received the fibrin-targeted contrast agent EP-2104R.

    Discussion

    In the present study, selective molecular MR imaging of pulmonary emboli and coronary thrombi was successfully demonstrated in vivo after systemic (intravenous) application of the novel fibrin-specific contrast agent EP-2104R. With this contrast agent, a high local gadolinium concentration in the clots could be achieved, allowing selective visualization of pulmonary emboli (n=32), intracardiac emboli (n=3), and coronary thrombi (n=5) as white spots, while the surrounding blood pool and soft tissue was signal suppressed. A new molecular MR contrast method for direct visualization of thrombus formation has extended former MR angiography techniques that have visualized the blood pool only.

    Over the past decade, central and lobal pulmonary emboli have been successfully visualized by MR imaging during the first pass of extracellular contrast medium after bolus contrast media injection (pulmonary MR angiography). With such a technique, the vessel lumen appears bright, whereas emboli can be depicted as filling defects.20,21 However, this method often fails to visualize more distal (subsegemental) emboli22 because data acquisition has to be performed within a single breathhold, resulting in a limited spatial resolution. In addition, because in these conventional MR angiography techniques the embolus and the lung tissue are signal suppressed, the detection of more distal filling defects or the lack of blood signal in smaller branching vessels as a result of occlusive emboli may be challenging. Motion errors from incomplete breathholding and intrinsic cardiac motion further reduce image quality.8,20,21 In contrast, with the present approach, direct and positive visualization of the emboli with high contrast to the blood pool and the surrounding tissue was established. Furthermore, navigator technology and cardiac triggering was used, allowing motion-artifact–free chest imaging with high spatial resolution. Patients with chest pain and dyspnea (ie, patients with pulmonary embolism or cardiac diseases) may benefit from such a free-breathing MR imaging method because in these patients reduced image quality is seen more often in standard contrast-enhanced breathhold MR angiography.20,21 Finally, molecular MR imaging can also be easily combined with MR venography21,23 for detection of deep vein thrombosis without further contrast media application.

    Other modalities for noninvasive imaging of pulmonary embolism may include spiral CT, which gives results similar to those of contrast-enhanced MRA. Spiral CT in conjunction with multidetector arrays is highly accurate for the detection of pulmonary emboli in animals (like pigs) and can be used as an alternative modality in these models.16,18 However, in patients, several pitfalls have been reported, and visualization of subsegmental thrombi may be limited.7,17,18,21 Furthermore, CT requires x-ray exposure and administration of potentially nephrotoxic contrast agent. There is still no single noninvasive imaging modality available with sufficient high negative predictive value to rule out pulmonary embolism in all patients, especially because isolated subsegmental embolism may occur in up to 30%.7

    Over the last decade, coronary MRA using a navigator-gated free-breathing cardiac-triggered T2-prepared 3D gradient-echo sequence has been implemented, allowing visualization of the vessel lumen with a bright signal while stenoses can be seen as lumen narrowing.24 However, with this technique, only the coronary lumen can be delineated; the composition of the stenoses (like soft plaque or thrombus formation) cannot be differentiated. Thrombus formation after rupture of a vulnerable plaque is known to be the main cause of myocardial infarction but is also seen in acute coronary syndromes like unstable angina or non–ST-elevation myocardial infarctions.1,3,4,6 In non–ST-elevated infarction, fibrin-rich or mixed thrombus layers are seen, and transmural infarction is typically associated with complete vessel occlusion by more red thrombus formation. Selective (targeted) MR imaging of coronary thrombosis may be an ideal noninvasive imaging modality for the visualization of all thrombi, including those associated with plaque rupture. In patients with acute chest pain and suspected myocardial ischemia without ECG changes (the benefit of invasive diagnosis with x-ray catheter in these patients is controversial6), direct molecular imaging of coronary thrombosis would be advantageous. In this study, a positive imaging test for noninvasive visualization of coronary thrombi was demonstrated through the use of molecular, fibrin-targeted MR imaging. Because of the high impact of thrombus formation in the progression of coronary artery disease and because thrombosis is a known cause of acute coronary syndromes,3,4 the present molecular thrombus MR imaging approach may have a very high clinical potential in patients with chest pain and suspected myocardial ischemia or with non–ST-elevation infarction. Furthermore, in these patients, less invasive imaging may be especially helpful and predictive because recent studies have suggested a substantial time delay between first thrombus formation and subsequent occlusion.5 Thus, early detection of nonocclusive coronary thrombus may lead to earlier intervention and prevention of acute myocardial infarction. In addition, other life-threatening thrombosis-related diseases in patients with acute chest pain such as pulmonary embolism may be depicted concurrently with coronary artery thrombosis. Both diseases are important differential diagnoses in patients with acute chest pain.1,2 Other indications for specific targeting of fibrin in MR imaging with high clinical impact may include deep vein thrombosis detection, clot visualization in patients with stroke, characterization of intraplaque hemorrhage,25,26 and cardiac thrombus visualization (ie, thrombus in the left atrial appendage). Also, molecular MR imaging of thrombus may enable a new tool for more detailed understanding of the impact of thrombus formation in the development of arteriosclerosis.

    Multilayer thrombi, consisting of platelet-rich thrombus and red thrombus, are often indicative of intermittent, clinically asymptomatic thrombus formation. These portions are likely related to distinct pathogenic mechanisms.27 Because each subsequent thrombus layer has the risk of causing myocardial infarction, old layers may be visualized by fibrin-specific MR imaging, which may allow risk stratification. However, because the amount of fibrin may vary in different thrombus layers,4 further investigations are needed to fully understand the impact of fibrin-binding molecular MR imaging in the field of arteriosclerosis.

    A small number of other thrombus-targeting agents such as nanoparticles and ultrasmall iron oxide particles have been developed.28,29 However, sufficient local concentration and successful chest MR imaging in a human size animal and after systemic (intravenous) contrast media administration have not been reported.15 Alternative MR imaging techniques of thrombus visualization may include direct clot imaging, which uses the T1 shortening of methemoglobin.30 However, such signal enhancement cannot be used in the first hours after clot formation, as is shown in our control swine in which clots could not be seen in the first 6 hours after clot delivery.31

    Limitations of the present study include the fact that only 1 dosage of the fibrin-targeted contrast EP-2104R was investigated. Furthermore, the optimal time point of MR imaging and the time curve of the Gd concentration in clots remain to be investigated. In this first study, only a model of acute clots was used. Old thrombi are also fibrin rich, but the efficacy of the new compound for detection of clots with different ages remains to be studied. Patient studies are needed to fully understand the potential of molecular thrombus imaging in clinical routine and to evaluate its clinical relevance in patients with acute chest pain.

    In conclusion, molecular MR imaging using the fibrin-targeted contrast-agent EP-2104R allows selective visualization of acute coronary, cardiac, and pulmonary thrombi.

    Acknowledgments

    This study was funded in part by EPIX Pharmaceuticals (Cambridge, Mass). We thank Richard Looby, Kirsten Overoye-Chan, Shirley Kalogeropoulos, and Steven Witowski for the synthesis and characterization of EP-2104R. We also thank Charles Chesna and Ronald Humphrey for assistance with the ICP samples.

    Footnotes

    Guest Editor for this article was Burton E. Sobel, MD.

    References

    Lee TH, Goldman L. Evaluation of the patient with acute chest pain. N Engl J Med. 2000; 342: 1187–1195.

    Kontos MC. Evaluation of the Emergency Department chest pain patient. Cardiol Rev. 2001; 9: 266–275.

    Falk E, Shah PK, Fuster V. Coronary plaque disruption. Circulation. 1995; 92: 657–671.

    Rentrop KP. Thrombi in acute coronary syndromes: revisited and revised. Circulation. 2000; 101: 1619–1626.

    Ojio S, Takatsu H, Tanaka T, Ueno K, Yokoya K, Matsubara T, Suzuki T, Watanabe S, Morita N, Kawasaki M, Nagano T, Nishio I, Sakai K, Nishigaki K, Takemura G, Noda T, Minatoguchi S, Fujiwara H. Considerable time from the onset of plaque rupture and/or thrombi until the onset of acute myocardial infarction in humans: coronary angiographic findings within 1 week before the onset of infarction. Circulation. 2000; 102: 2063–2069.

    Prasad A, Mathew V, Holmes DR Jr, Gersh BJ. Current management of non-ST-segment-elevation acute coronary syndrome: reconciling the results of randomized controlled trials. Eur Heart J. 2003; 24: 1544–1553.

    Fedullo PF, Tapson VF. Clinical practice: the evaluation of suspected pulmonary embolism. N Engl J Med. 2003; 349: 1247–1256.

    Olin JW. Pulmonary embolism. Rev Cardiovasc Med. 2002; 3: S68–S75.

    Miniati M, Prediletto R, Formichi B, Marini C, Di Ricco G, Tonelli L, Allescia G, Pistolesi M. Accuracy of clinical assessment in the diagnosis of pulmonary embolism. Am J Respir Crit Care Med. 1999; 159: 864–871.

    Giannitsis E, Muller-Bardorff M, Kurowski V, Weidtmann B, Wiegand U, Kampmann M, Katus HA. Independent prognostic value of cardiac troponin T in patients with confirmed pulmonary embolism. Circulation. 2000; 102: 211–217.

    Fleckenstein JL, Archer BT, Barker BA, Vaughan JT, Parkey RW, Peshock RM. Fast short-tau inversion-recovery MR imaging. Radiology. 1991; 179: 499–504.

    Spuentrup E, Buecker A, Stuber M, Botnar R, Nguyen TH, Bornert P, Kolker C, Gunther RW. Navigator-gated coronary magnetic resonance angiography using steady-state-free-precession: comparison to standard t2-prepared gradient-echo and spiral imaging. Invest Radiol. 2003; 38: 263–268.

    Stuber M, Botnar RM, Spuentrup E, Kissinger KV, Manning WJ. Three-dimensional high-resolution fast spin echo coronary magnetic resonance angiography. Magn Reson Med. 2001; 45: 206–211.

    Botnar RM, Perez AS, Witte S, Wiethoff AJ, Laredo J, Hamilton J, Quist W, Parsons EC Jr, Vaidya A, Kolodziej A, Barrett JA, Graham PB, Weisskoff RM, Manning WJ, Johnstone MT. In vivo molecular imaging of acute and subacute thrombosis using a fibrin-binding magnetic resonance imaging contrast agent. Circulation. 2004; 109: 2023–2029.

    Botnar R, Buecker A, Wiethoff AJ, Parsons EC Jr, Katoh M, Katsimaglis G, Weisskoff RM, Lauffer RB, Graham PB, Gunther RW, Manning WJ, Spuentrup E. In vivo magnetic resonance imaging of coronary thrombosis using a fibrin-binding molecular magnetic resonance contrast agent. Circulation. 2004; 110: 1463–1466.

    Baile EM, King GG, Muller NL, D’Yachkova Y, Coche EE, Pare PD, Mayo JR. Spiral computed tomography is comparable to angiography for the diagnosis of pulmonary embolism. Am J Respir Crit Care Med. 2000; 161: 1010–1015.

    Ghaye B, Szapiro D, Mastora I, Delannoy V, Duhamel A, Remy J, Remy-Jardin M. Peripheral pulmonary arteries: how far in the lung does multi-detector row spiral CT allow analysis; Radiology. 2001; 219: 629–636.

    Remy-Jardin M, Mastora I, Remy J. Pulmonary embolus imaging with multislice CT. Radiol Clin North Am. 2003; 41: 507–519.

    Ghaye B, Remy J, Remy-Jardin M. Non-traumatic thoracic emergencies: CT diagnosis of acute pulmonary embolism: the first 10 years. Eur Radiol. 2002; 12: 1886–1905.

    Meaney JF, Weg JG, Chenevert TL, Stafford-Johnson D, Hamilton BH, Prince MR. Diagnosis of pulmonary embolism with magnetic resonance angiography. N Engl J Med. 1997; 336: 1422–1427.

    Tapson VF. Pulmonary embolism: new diagnostic approaches. N Engl J Med. 1997; 336: 1449–1451.

    Gupta A, Frazer CK, Ferguson JM, Kumar AB, Davis SJ, Fallon MJ, Morris IT, Drury PJ, Cala LA. Acute pulmonary embolism: diagnosis with MR angiography. Radiology. 1999; 210: 353–359.

    Evans AJ, Sostman HD, Knelson MH, Spritzer CE, Newman GE, Paine SS, Beam CA. 1992 ARRS Executive Council Award: detection of deep venous thrombosis: prospective comparison of MR imaging with contrast venography. AJR Am J Roentgenol. 1993; 161: 131–139.

    Kim WY, Danias PG, Stuber M, Flamm SD, Plein S, Nagel E, Langerak SE, Weber OM, Pedersen EM, Schmidt M, Botnar RM, Manning WJ. Coronary magnetic resonance angiography for the detection of coronary stenoses. N Engl J Med. 2001; 345: 1863–1869.

    Corti R, Fuster V, Badimon JJ. Pathogenetic concepts of acute coronary syndromes. J Am Coll Cardiol. 2003; 41: 7S–14S.

    Kramer CM. Magnetic resonance imaging to identify the high-risk plaque. Am J Cardiol. 2002; 90: 15L–17L.

    Maseri A, Fuster V. Is there a vulnerable plaque; Circulation. 2003; 107: 2068–2071.

    Flacke S, Fischer S, Scott MJ, Fuhrhop RJ, Allen JS, McLean M, Winter P, Sicard GA, Gaffney PJ, Wickline SA, Lanza GM. Novel MRI contrast agent for molecular imaging of fibrin: implications for detecting vulnerable plaques. Circulation. 2001; 104: 1280–1285.

    Yu X, Song SK, Chen J, Scott MJ, Fuhrhop RJ, Hall CS, Gaffney PJ, Wickline SA, Lanza GM. High-resolution MRI characterization of human thrombus using a novel fibrin-targeted paramagnetic nanoparticle contrast agent. Magn Reson Med. 2000; 44: 867–872.

    Moody AR, Murphy RE, Morgan PS, Martel AL, Delay GS, Allder S, MacSweeney ST, Tennant WG, Gladman J, Lowe J, Hunt BJ. Characterization of complicated carotid plaque with magnetic resonance direct thrombus imaging in patients with cerebral ischemia. Circulation. 2003; 107: 3047–3052.

    Corti R, Osende JI, Fayad ZA, Fallon JT, Fuster V, Mizsei G, Dickstein E, Drayer B, Badimon JJ. In vivo noninvasive detection and age definition of arterial thrombus by MRI. J Am Coll Cardiol. 2002; 39: 1366–1373.(Elmar Spuentrup, MD; Arno)