     Precision of abutment tooth model reconstructed by 3D laser scanner

    HAN YanFeng, GAO Bo, WU Jiang, HU Jiang

    Department of Prosthodontics, Stomatological College, Fourth Military Medical University, Xian 710033, China

    【Abstract】 AIM: To evaluate the precision of the threedimensional (3D) laser scanner for scanning abutment tooth and reconstructing its model so as to offer some evidence for its application in CAD/CAM of oral prosthesis. METHODS: Five plaster supramaxillary abutment tooth models were scanned and reconstructed by 3D laser scanner. The 3D shape of the models was measured by Surfacer10.5 software and the plaster models were measured manually with vernier caliper. The data were statistically analyzed. RESULTS: The 3D laser scanner showed the better reliability and repeatability (r>0.75, P<0.05). There were no significant differences between the data acquired by scanning and manual measurement in the precision (P<0.05). CONCLUSION: 3D laser scanner is precise and reliable enough for reconstructing the abutment tooth model fit for the clinical application. ......