Effect of Nasal Continuous Positive Airway Pressure Treatment on Blood Pressure in Patients With Obstructive Sleep Apnea
http://www.100md.com 《循环学杂志》2003年第1期
     From the Department of Respiratory Medicine, Philipps-University Marburg, Marburg, Germany, and the Department of Medicine, David Read Laboratory, University of Sydney, Australia (C.E.S.).$2;^}9/, 百拇医药

    Abstract$2;^}9/, 百拇医药

    Background— There is increasing evidence that obstructive sleep apnea (OSA) is an independent risk factor for arterial hypertension. Because there are no controlled studies showing a substantial effect of nasal continuous positive airway pressure (nCPAP) therapy on hypertension in OSA, the impact of treatment on cardiovascular sequelae has been questioned altogether. Therefore, we studied the effect of nCPAP on arterial hypertension in patients with OSA.$2;^}9/, 百拇医药

    Methods and Results— Sixty consecutive patients with moderate to severe OSA were randomly assigned to either effective or subtherapeutic nCPAP for 9 weeks on average. Nocturnal polysomnography and continuous noninvasive blood pressure recording for 19 hours was performed before and with treatment. Thirty two patients, 16 in each group, completed the study. Apneas and hypopneas were reduced by 95% and 50% in the therapeutic and subtherapeutic groups, respectively. Mean arterial blood pressure decreased by 9.9±11.4 mm Hg with effective nCPAP treatment, whereas no relevant change occurred with subtherapeutic nCPAP (P=0.01). Mean, diastolic, and systolic blood pressures all decreased significantly by 10 mm Hg, both at night and during the day.

    Conclusions— Effective nCPAP treatment in patients with moderate to severe OSA leads to a substantial reduction in both day and night arterial blood pressure. The fact that a 50% reduction in the apnea-hypopnea index did not result in a decrease in blood pressure emphasizes the importance of highly effective treatment. The drop in mean blood pressure by 10 mm Hg would be predicted to reduce coronary heart disease event risk by 37% and stroke risk by 56%./gu, 百拇医药

    Key Words: hypertension cardiovascular diseases sleep blood pressure/gu, 百拇医药

    Introduction/gu, 百拇医药

    In recent years it has been shown that obstructive sleep apnea (OSA) is a common disorder in adults.1 OSA is characterized by repetitive apneas and hypoxia caused by upper airway collapse during sleep, despite respiratory efforts of the diaphragm. Five or more apneas per hour of sleep are generally considered abnormal, and severely affected patients have several hundred apneas each night. Most apneas and hypopneas are terminated by a transient arousal from sleep and consecutive hyperventilation. Disruption of normal sleep by frequent arousals leads to excessive daytime sleepiness, the most prominent symptom in these patients.

    Nocturnal arterial blood pressure is increased in OSA patients, and there is increasing evidence that OSA is an independent risk factor for arterial hypertension during the day.2–5 Although the exact mechanisms are still unclear, a persistent increase in sympathetic tone caused by chronically occurring repetitive hypoxia and arousal are thought to be the key mechanisms for the short- and long-term blood pressure increases in OSA.6,7 It has also recently been shown that patients with OSA have an impairment of resistance-vessel endothelium-dependent vasodilation.8ty3\+%, http://www.100md.com

    Nasal continuous positive airway pressure (nCPAP) has become the standard treatment for OSA9 and has been shown in controlled studies to reduce symptoms and improve quality of life in OSA patients.10,11 Controlled studies showed either no effect12,13 or only a minor decrease in arterial blood pressure by 1.4 and 2.5 mm Hg, respectively.14,15 The effectiveness of this treatment on cardiovascular sequelae in OSA patients has been questioned altogether.16 Therefore, we performed a prospective randomized trial to evaluate the effect of nCPAP on arterial blood pressure in OSA patients.

    Methods/io, http://www.100md.com

    Patients and Protocol/io, http://www.100md.com

    The study was conducted in the Sleep Unit at the University Hospital in Marburg, Germany. The Unit takes adult patients referred by general practitioners (35%), pneumologists (35%), or other subspecialties (30%). During the study period, 283 patients fulfilled the inclusion criteria of 5 apneas or hypopneas per hour of sleep and excessive daytime sleepiness (10 or more points out of a maximum of 24 on the Epworth sleepiness scale17;). A maximum of 4 patients per week could be included in the study. If there was more than one patient eligible on one day, the patient with the most pronounced sleep apnea according to an ambulatory recording was asked first to participate in the study./io, http://www.100md.com

    fig.ommitted/io, http://www.100md.com

     Study profile. EDS indicates excessive daytime sleepiness./io, http://www.100md.com

    Exclusion criteria were predominantly central sleep apnea, respiratory failure, heart failure (NYHA class III or IV), myocardial infarction 3 months before the study, and relevant cardiac arrhythmia (second- and third-degree heart block or premature ventricular contractions in Lown classes IV or V). Professional drivers were also excluded. The local ethics committee approved the study, and all subjects gave written informed consent.

    Sixty consecutive patients who consented to participate underwent baseline diagnostic polysomnography and continuous noninvasive blood pressure recording for 19.1±1.3 hours using the Portapres device (TNO Biomedical Instrumentation)..jlv*, 百拇医药

    Patients were then randomized to receive either effective or subtherapeutic nCPAP treatment. Randomization was performed on the telephone by a person who was otherwise not involved in the study..jlv*, 百拇医药

    All study patients spent another 2 nights in the sleep laboratory to ensure adaptation to the nasal mask. Treatment was done with either the Sullivan IV (ResMed) or the Aria (Respironics) nCPAP device. In the effective treatment group, treatment pressure was increased until apneas, hypopneas, and snoring were prevented during all sleep stages and with the patient lying supine. Effective treatment pressure was 9.1±2.3 cm H2O on average (range, 6 to 12 cm H2O). In the subtherapeutic treatment group, pressure was left unchanged at the lowest possible value for the nCPAP device used (3 or 4 cm H2O, respectively).

    Patients then went home on the prescribed treatment for an average of 65.2±49.6 days. They were then readmitted to the sleep laboratory, and polysomnography and continuous blood pressure measurements during daytime and sleep were repeated on either subtherapeutic or effective nCPAP treatment. Compliance was objectively measured by the built-in compliance software of the nCPAP devices.\e!t, http://www.100md.com

    Any antihypertensive medication the patients were taking at the beginning of the study remained unchanged throughout the entire study. Patients were considered dropouts if antihypertensive medication was accidentally changed by the patient or their general practitioner.\e!t, http://www.100md.com

    In 28 patients the study could not be completed because of technical problems with the blood pressure device (n=11), a change in antihypertensive treatment (n=7, 3 in the effective and 4 in the subtherapeutic nCPAP group), or the patient did not want to continue the study (n=8) or did not tolerate treatment (n=2).

    Complete measurements were available in 32 patients, 16 on effective and 16 on subtherapeutic nCPAP. These data are reported here. Although arterial hypertension was neither an inclusion nor an exclusion criterion, 8 of 16 patients in the effective group and 13 of 16 patients in the subtherapeutic group turned out to be hypertensive (15 of these patients were on antihypertensive treatment and 6 had an office blood pressure of 160 and/or 90 mm Hg or more). Seven of the 15 patients on antihypertensive medication were treated with one drug (ACE inhibitor, n=4; angiotensin II receptor blocker, n=2; and calcium channel blocker, n=1). A combination of 2 antihypertensive drugs was used in 7 patients (angiotensin II receptor blocker plus diuretic, n=4; ACE inhibitor plus diuretic, n=1; ACE inhibitor plus ß-blocker, n=1; and calcium channel blocker plus diuretic, n=1), and one patient was on a combination of 4 antihypertensives (ACE inhibitor, diuretic, calcium channel blocker, and {alpha} -blocker).

    Anthropometric data and relevant diagnoses are shown in*:r^%], http://www.100md.com

    fig.ommitted*:r^%], http://www.100md.com

     Anthropometric Data and Relevant Diagnoses*:r^%], http://www.100md.com

    Measurements*:r^%], http://www.100md.com

    Sleep Study Recordings*:r^%], http://www.100md.com

    Polysomnography was performed and visually scored according to standard criteria,18,19 as detailed previously.20 The average number of apneas and hypopneas per hour of sleep (apnea-hypopnea index [AHI]) was calculated.*:r^%], http://www.100md.com

    Blood Pressure Recording*:r^%], http://www.100md.com

    The Portapres is a battery-operated portable instrument to measure arterial blood pressure continuously and noninvasively with 2 finger cuffs.21,22 Portapres provides a height correction to compensate for hydrostatic level effects due to movements of the measured finger with respect to the reference point at heart level. Blood pressure values measured by Portapres are in good concordance with invasively measured values21,23,24 and are highly reproducible.24,25 Because of the battery we were using, recording time was limited to 20 hours. After the elimination of artifacts, an average of 19.1±1.3 hours of blood pressure recordings were analyzed. During the diagnostic sleep study, patients spent 7.2±1 hours in bed and 7.1±1 hours on treatment. Nighttime blood pressure was calculated for this time period. Daytime blood pressure was thus recorded during the remaining 12 hours.

    Patients were allowed to move freely in the hospital, but they were not able to climb stairs because the equipment was mounted on an infusion pole.q*zh'[m, 百拇医药

    After the recording, blood pressure data were processed automatically with the software of the device, including the built-in artifact removal. No further processing of the data was done by the investigators. For each minute of the recording, the software of the device calculated mean, systolic, and diastolic blood pressure as well as heart rate.q*zh'[m, 百拇医药

    Statisticsq*zh'[m, 百拇医药

    Mean arterial blood pressure during the entire recording period was the primary outcome variable. Secondary outcome variables were systolic and diastolic blood pressure during the entire recording period, as well as daytime and nighttime mean, systolic, and diastolic blood pressures. Tertiary outcome variables were the changes in polysomnographic parameters (AHI, sleep stages, SaO2) and sleepiness (Epworth sleepiness scale).

    Statistical analysis was performed using the SPSS statistical package, version 10.0 (SPSS). Two-factor ANOVA (group versus time) with repeated measures on the factor time (baseline minus treatment) was used to test the effect of therapeutic versus subtherapeutic nCPAP. To test for differences between the 2 treatment groups at baseline and for differences in nCPAP compliance, the unpaired t test (2 sided) was used. Data are reported as mean±SD unless otherwise stated. Statistical significance was assumed at P0.05.s;, http://www.100md.com

    Resultss;, http://www.100md.com

    There was a decrease in mean arterial blood pressure over the 19.1±1.3-hour recording period by 9.9±11.4 mm Hg in the therapeutic nCPAP group, but an increase by 0.6±10.8 mm Hg in the subtherapeutic nCPAP group (P=0.01, ANOVA interaction time by group). Both diastolic and systolic blood pressures also decreased significantly by 10.3±11.4 mm Hg and 9.5±15.0 mm Hg, respectively, with therapeutic nCPAP (P<0.005 and P=0.04, respectively, ANOVA interaction time by group) compared with subtherapeutic nCPAP. The decrease in mean blood pressure with effective nCPAP was present both during the day (-10.0±12.1 mm Hg) and night (-10.3±15.3 mm Hg). These effects of subtherapeutic and therapeutic nCPAP on blood pressure are summarized in The time course of mean arterial blood pressure before and on treatment for both groups is shown in . In the effective treatment group, mean arterial blood pressure decreased during the entire recording period; the most pronounced drop occurred during the night and in the morning until about noon. In the afternoon and evening, the blood pressure decrease was still present, but less pronounced, when compared with the night and the first part of the day.

    fig.ommitted8yx5(, http://www.100md.com

     Changes in blood pressure with effective (closed bars) and subtherapeutic (open bars) nCPAP. Significant difference. MAP indicates mean arterial blood pressure; systolic, systolic blood pressure; and diastolic, diastolic blood pressure. MAP, P=0.01; systolic blood pressure, P=0.04; diastolic blood pressure, P<0.005.8yx5(, http://www.100md.com

    fig.ommitted8yx5(, http://www.100md.com

     Blood Pressure (mm Hg) at Baseline and on Treatment8yx5(, http://www.100md.com

    fig.ommitted8yx5(, http://www.100md.com

     A, Time course of mean arterial blood pressure (MAP) before (closed circles) and on treatment (open circles) with therapeutic nCPAP. On average, 7.2±1 hours were recorded during the night. B, Time course of mean arterial blood pressure (MAP) before (closed circles) and on treatment (open circles) with subtherapeutic nCPAP. On average, 7.1±1 hours were recorded during the night.8yx5(, http://www.100md.com

    The results of polysomnography before and on treatment are presented in There were no significant differences in any of the variables between the 2 groups at baseline. Except for total sleep time, all parameters improved with both treatment modalities. The improvements of AHI, sleepiness, and mean SaO2 were significantly larger with therapeutic nCPAP compared with subtherapeutic nCPAP (ANOVA interaction time by group).

    fig.ommittedn?5[kh*, http://www.100md.com

     Polysomnography at Baseline and on Treatmentn?5[kh*, http://www.100md.com

    Compliance with nCPAP was high in both groups: the average use per night was 5.5±2.0 hours in the therapeutic and 5.4±2.2 hours in the subtherapeutic group (P=NS). Body weight was similar in both groups at baseline (103.1±16.5 kg and 102.6±17.8 kg in the therapeutic and subtherapeutic groups, respectively; P=NS), and it remained virtually unchanged (103.0±16.0 kg and 102.3±17.1 kg in the therapeutic and subtherapeutic groups, respectively; P=NS, ANOVA interaction time by group).n?5[kh*, http://www.100md.com

    Discussionn?5[kh*, http://www.100md.com

    After an average of 65 days of treatment, we found a decrease in mean arterial blood pressure by 9.9 mm Hg with effective nCPAP, whereas there was a slight increase in blood pressure with subtherapeutic nCPAP. Mean, diastolic, and systolic blood pressures decreased significantly during both the day and night by between 8.1 and 11.4 mm Hg. Although subtherapeutic nCPAP led to a significant reduction of AHI by 50%, blood pressure did not decrease.

    There is general agreement that nCPAP treatment leads to a reversal of apnea-associated short-term blood pressure increases during sleep26,27; however, the effect on daytime blood pressure is uncertain. Whereas 3 uncontrolled studies found a 24-hour blood pressure decrease with nCPAP,25,27,28 4 controlled studies found no significant decrease12,13 or only a minor blood pressure reduction (1.4 and 2.5 mm Hg, respectively).14,15n!{d, 百拇医药

    In contrast to these previously published controlled studies, our data show a pronounced reduction in day and night blood pressure with therapeutic nCPAP. Several factors may explain the discrepancy in the results. (1) The different results may have been caused by the fact that treatment duration was substantially longer in our study and that the use of nCPAP was high in our patients. (2) A total of 21 of our 32 patients were hypertensive, because mainly patients with moderate to severe OSA were included, whereas hypertensives were excluded in one study14 or less prevalent in the others.12,13 (3) The different methods of blood pressure measurement will influence the results. It has been shown that 24-hour ambulatory blood pressure measurement causes arousal from sleep in 64% of the recordings and leads to an increase in systolic and diastolic blood pressure by 13.7±15.9 mm Hg and 3.7±8.2 mm Hg, respectively.29 The ambulatory blood pressure measurement will therefore underestimate the changes in nighttime blood pressure caused by nCPAP. An important advantage of Portapres is that it does not cause arousal from sleep.30 Furthermore, continuous blood pressure data, as measured in our study, are in good concordance with invasively measured blood pressures, are highly reproducible,21 and therefore more accurately mirror the actual blood pressure than intermittent measurements.

    There was no relevant change in body weight in either group, so this factor did not contribute to our results.1'uw9, 百拇医药

    The increase in sympathetic activity caused by apnea-associated repetitive asphyxia and arousal are thought to be the key mechanisms in the pathogenesis of daytime arterial hypertension in patients with OSA. Treatment with nCPAP attenuates the increase in sympathetic activity.31 In an animal model, repetitive obstructive apneas and acoustic arousal from sleep led to a similar increase in nighttime blood pressure. However, recurrent arousal from sleep without airway obstruction did not lead to an increase in daytime blood pressure, whereas sustained hypertension developed if apneas had been produced in the animals.32 Although most polysomnographic parameters improved more markedly with effective compared with subtherapeutic nCPAP, apart from the AHI and Epworth sleepiness scale scores, only mean SaO2 was significantly higher on therapeutic than on subtherapeutic nCPAP. This finding further emphasizes the role of hypoxia in the evolution of daytime arterial hypertension and the importance of preventing hypoxia to achieve optimal treatment of arterial hypertension in OSA patients.

    In our study there was a substantial number of drop outs, mainly because of technical problems and changes in antihypertensive medication. Portapres is a complex device and is therefore more prone to technical problems than ambulatory blood pressure measurement. Because technical defects and accidental changes in medication occurred in both groups in a similar number of patients, this should not have influenced our results.!dnoms, 百拇医药

    The reduced number of patients that could be included in the final analysis has reduced power from the statistical tests performed; however, because of the large effect of therapeutic nCPAP on blood pressure, differences in primary and secondary outcome variables were significant.!dnoms, 百拇医药

    We used a single-blind study design because a method for applying therapeutic and subtherapeutic nCPAP in a double-blind fashion was not available. The knowledge of the allocation to one or the other treatment group might have influenced the analysis of our data. Polysomnography was therefore scored by a technician who was not informed about the study, and blood pressure analysis, including artifact recognition and elimination, was done exclusively by the software of the Portapres device. The risk of a bias due to the single-blind design was thus minimized.

    Unexpectedly, nCPAP at a pressure of 3 or 4 cm H2O reduced mean AHI by 50%, improved sleep structure, and reduced desaturation. NCPAP at the pressure applied is therefore not a placebo but a suboptimal form of treatment, because even the low treatment pressure used here may be sufficient to at least partly reverse upper airway obstruction in many patients. However, the reduction in AHI in the control group would have acted against our hypothesis that nCPAP lowers blood pressure. Despite the reduction in AHI in the subtherapeutic treatment group by 50%, there was no reduction in blood pressure in this group. The unexpected result that suboptimal nCPAP has a substantial effect on AHI but no effect on blood pressure emphasizes the importance of optimal treatment to reduce cardiovascular sequelae.)0, 百拇医药

    This is the first prospective, randomized study to demonstrate a substantial reduction in arterial blood pressure during both the day and night with 9 weeks of therapeutic nCPAP treatment compared with subtherapeutic nCPAP. Although AHI was reduced by 50% with subtherapeutic nCPAP, there was no effect on blood pressure, emphasizing the importance of optimal treatment. In our study, the drop in mean blood pressure was close to 10 mm Hg. This would be predicted to reduce coronary heart disease event risk by 37% and stroke risk by 56%.33

    Acknowledgmentsqg), http://www.100md.com

    Supported in part by grants from ResMed Germany (Mönchengladbach, Germany) and Heinen & Löwenstein (Bad Ems, Germany). We thank Wilma Althaus for technical assistance with the scoring of the sleep studies and Werner Cassel for his assistance with the statistical analysis. We thank Dr John Stradling for his valuable comments and discussion of the manuscript.qg), http://www.100md.com

    Referencesqg), http://www.100md.com

    Young T, Palta M, Dempsey J, et al. The occurrence of sleep-disordered breathing among middle-aged adults. N Engl J Med. 1993; 328: 1230–1235.qg), http://www.100md.com

    Grote L, Ploch T, Heitmann J, et al. Sleep-related breathing disorder is an independent risk factor for systemic hypertension. Am J Respir Crit Care Med. 1999; 160: 1875–1882.qg), http://www.100md.com

    Peppard PE, Young T, Palta M, et al. Prospective study of the association between sleep-disordered breathing and hypertension. N Engl J Med. 2000; 342: 1378–1384.qg), http://www.100md.com

    Nieto FJ, Young TB, Lind BK, et al. Association of sleep-disordered breathing, sleep apnea, and hypertension in a large community-based study: Sleep Heart Health Study. JAMA. 2000; 283: 1829–1836.

    Lavie P, Herer P, Hoffstein V. Obstructive sleep apnoea syndrome as a risk factor for hypertension: population study. BMJ. 2000; 320: 479–482.fh^ej, 百拇医药

    Carlson JT, Hedner J, Elam M, et al. Augmented resting sympathetic activity in awake patients with obstructive sleep apnea. Chest. 1993; 103: 1763–1768.fh^ej, 百拇医药

    Narkiewicz K, van de Borne PJ, Cooley RL, et al. Sympathetic activity in obese subjects with and without obstructive sleep apnea. Circulation. 1998; 98: 772–776.fh^ej, 百拇医药

    Kato M, Roberts-Thomson P, Phillips BG, et al. Impairment of endothelium-dependent vasodilation of resistance vessels in patients with obstructive sleep apnea. Circulation. 2000; 102: 2607–2610.fh^ej, 百拇医药

    Sullivan CE, Issa FG, Berthon-Jones M, et al. Reversal of obstructive sleep apnoea by continuous positive airway pressure applied through the nares. Lancet. 1981; 1: 862–865.fh^ej, 百拇医药

    Engleman HM, Martin SE, Deary IJ, et al. Effect of continuous positive airway pressure treatment on daytime function in sleep apnoea/hypopnoea syndrome. Lancet. 1994; 343: 572–575.

    Jenkinson C, Davies RJ, Mullins R, et al. Comparison of therapeutic and subtherapeutic nasal continuous positive airway pressure for obstructive sleep apnoea: a randomised prospective parallel trial. Lancet. 1999; 353: 2100–2105.r{, 百拇医药

    Dimsdale JE, Loredo JS, Profant J. Effect of continuous positive airway pressure on blood pressure: a placebo trial. Hypertension. 2000; 35: 144–147.r{, 百拇医药

    Engleman HM, Gough K, Martin SE, et al. Ambulatory blood pressure on and off continuous positive airway pressure therapy for the sleep apnea/hypopnea syndrome: effects in "non-dippers". Sleep. 1996; 19: 378–381.r{, 百拇医药

    Faccenda JF, Mackay TW, Boon NA, et al. Randomized placebo-controlled trial of continuous positive airway pressure on blood pressure in the sleep apnea-hypopnea syndrome. Am J Respir Crit Care Med. 2001; 163: 344–348.r{, 百拇医药

    Pepperell JC, Ramdassingh-Dow S, et al. Ambulatory blood pressure after therapeutic and subtherapeutic nasal continuous positive airway pressure for obstructive sleep apnoea: a randomised parallel trial. Lancet. 2002; 359: 204–210.

    Wright J, Johns R, Watt I, et al. Health effects of obstructive sleep apnoea and the effectiveness of continuous positive airways pressure: a systematic review of the research evidence. BMJ. 1997; 314: 851–860.;p+n:, http://www.100md.com

    Johns MW. Daytime sleepiness, snoring, and obstructive sleep apnea: the Epworth sleepiness scale. Chest. 1993; 103: 30–36.;p+n:, http://www.100md.com

    Rechtschaffen A, Kales A. A Manual of Standardized Terminology: Techniques and Scoring System for Sleep Stages of Human Subjects. Los Angeles: UCLA Brain Information Service/Brain Research Institute; 1968.;p+n:, http://www.100md.com

    American Thoracic Society. Indications and standards for cardiopulmonary sleep studies. Am Rev Respir Dis. 1989; 139: 559–568.;p+n:, http://www.100md.com

    Becker HF, Piper AJ, Flynn WE, et al. Breathing during sleep in patients with nocturnal desaturation. Am J Respir Crit Care Med. 1999; 159: 112–118.;p+n:, http://www.100md.com

    Imholz BP, Langewouters GJ, van Montfrans GA, et al. Feasibility of ambulatory, continuous 24-hour finger arterial pressure recording. Hypertension. 1993; 21: 65–73.;p+n:, http://www.100md.com

    Bos WJ, van Goudoever J, van Montfrans GA, et al. Reconstruction of brachial artery pressure from noninvasive finger pressure measurements. Circulation. 1996; 94: 1870–1875.

    Hirschl MM, Binder M, Herkner H, et al. Accuracy and reliability of noninvasive continuous finger blood pressure measurement in critically ill patients. Crit Care Med. 1996; 24: 1684–1689.pi, 百拇医药

    Voogel AJ, van Montfrans GA. Reproducibility of twenty-four-hour finger arterial blood pressure, variability and systemic hemodynamics. J Hypertens. 1997; 15: 1761–1765.pi, 百拇医药

    Voogel AJ, van Steenwijk RP, Karemaker JM at al. Effects of treatment of obstructive sleep apnea on circadian hemodynamics. J Auton Nerv Syst. 1999; 77: 177–183.pi, 百拇医药

    Mayer J, Becker H, Brandenburg U, et al. Blood pressure and sleep apnea: results of long-term nasal continuous positive airway pressure therapy. Cardiology. 1991; 79: 84–92.pi, 百拇医药

    Jennum P, Wildschiodtz G, Christensen NJ, et al. Blood pressure, catecholamines, and pancreatic polypeptide in obstructive sleep apnea with and without nasal continuous positive airway pressure (nCPAP) treatment. Am J Hypertens. 1989; 2: 847–852.pi, 百拇医药

    Wilcox I, Grunstein RR, Hedner JA, et al. Effect of nasal continuous positive airway pressure during sleep on 24-hour blood pressure in obstructive sleep apnea. Sleep. 1993; 16: 539–544.

    Heude E, Bourgin P, Feigel P, et al. Ambulatory monitoring of blood pressure disturbs sleep and raises systolic pressure at night in patients suspected of suffering from sleep-disordered breathing. Clin Sci (Colch). 1996; 91: 45–50.r$-&'f, 百拇医药

    Jerrentrup A, Becker HF, Heitmann J, et al. Does continuous non-invasive blood pressure recording affect sleep quality in patients with nCPAP-therapy? Eur Respir J. 1999; 14: 85s. Abstract.r$-&'f, 百拇医药

    Somers VK, Dyken ME, Clary MP, et al. Sympathetic neural mechanisms in obstructive sleep apnea. J Clin Invest. 1995; 96: 1897–1904.r$-&'f, 百拇医药

    Brooks D, Horner RL, Kozar LF, et al. Obstructive sleep apnea as a cause of systemic hypertension. Evidence from a canine model. J Clin Invest. 1997; 99: 106–109.r$-&'f, 百拇医药

    Macmahon S, Peto R, Cutler J, et al. Blood pressure, stroke, and coronary heart disease. part 1: prolonged differences in blood pressure: prospective observational studies corrected for the regression dilution bias. Lancet. 1990; 335: 765–774.(Heinrich F. Becker MD Andreas Jerrentrup MD Thomas Ploch Dipl Psych Ludger Grote, MD Thomas Penzel P)