GE公司Lightspeed故障2例
故障,1故障一,2故障二
GE公司Lightspeed故障2例 (pdf)随着医院高新大型设备的不断引进,医院设备维修越来越困难,维修费用也越来越高,这就需要工程技术人员在加强维修技术的培训和学习外,更要加强同行之间的信息沟通,本文提供有关GE Lightspeed机器故障2例,以供同仁参考。
1 故障一
1.1 故障现象 Lightspeed Ultra在使用过程中突然停止扫描,故障提示如下。
Suite:HSA1 Host:OBC Proc:Generator Error:183121
File:gen_abort.c 1.30 Method:No Method Line:340
Function:XRAY GENERATION:Filament Selection
Scan:0/0/0 Type:None/Unknown
Exception Level:Pri/Most Severe Time:17:02:22:885 Log Series:26
FIL/MA fault detected:Open filament circuit.
Desired focal spot=0.(0=large,1=small)
Fault latch address=ffdf43 hex,Bit=D7
Suite:HSA1 Host:OBC Proc:Generator Error:183121
File:gen_abort.c 1.30 Method:No Method Line:117
Function:XRAY GENERATION:Filament Selection
Scan:0/0/0 Type:None/Unknown
Exception Level:Pri/Most Severe Time:17:08:30:995 Log Series:28
FIL/MA fault detected:Open filament circuit.
Desired focal spot=1.(0=large,1=small)
Fault latch address=ffdf43 hex,Bit=D7
1.2 故障分析 OBC控制盒中毫安控制板调整球管灯丝板上灯丝逆变部分的输出电压 ......